Turbidity Sensor Spikes: Check Bubbles, Fouling and Real Events

Turbidity sensor spikes can reflect a genuine change in suspended matter or interference at the sensing surface. Check the timing against flow, rainfall, pumping and service events. Compare the sensor with a suitable independent observation before correcting the record or changing the calibration.
Preserve the original trace
Save the raw values, sample interval and any status flags. Check whether the spike exists in the instrument record as well as the dashboard. A displayed value or isolated outlier is a clue. Keep the unprocessed reading so it can be compared with what was happening at the sampling point.
Compare with a plausible change in the water
An increase after rainfall, a valve operation or disturbance near a sampling point may represent a real change. A signal from another nearby measurement can help establish timing. The USDA turbidity sampling guide describes field conditions that can produce both real and spurious excursions. The turbidity measurement guide explains the measurement principle.

Look for bubbles when the flow state changes
Inspect the sensing region or flow cell under the equipment’s approved procedure. Air bubbles in the sample or at the optical window can affect light scattering. If spikes recur during pump starts or intermittent sample supply, record that timing and look for direct evidence of bubbles. The NEMI method summary lists bubbles, fouling and settling among interferences.
Inspect the window and wiper
Deposits, algae, scratches or an obstructed wiper can change the reading. Inspect the optical surface and verify wiper operation using the model instructions. Save readings before and after cleaning where feasible. The USGS continuous monitor guidance discusses inspecting the window and cleaning system before calibration checks.
Compare measurements on the same basis
If practical, compare with a suitable instrument or sample taken at the same location and time. Handle the sample consistently to limit settling and air entrainment. Use the same method where possible. See the NTU and FNU comparison for why different optical methods may disagree on natural samples.
Document the cause and verify the next event
Record the suspected cause, observed conditions, maintenance time and verification result. When the evidence is incomplete, flag the affected interval and preserve the original values. Confirm normal operation through a comparable flow event when possible. If the location repeatedly traps air or debris, use the turbidity sensor installation guide to review the mounting and sample path.
Frequently asked questions
Should every isolated spike be removed? Use the project’s quality rules and retain evidence. A short event can still be real.
Will recalibration fix a bubbly sample? Check the sample path and window first. A calibration check verifies the response under its test conditions.